User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: PM 2522A Calibration (Ver1.4) DATE: 03-Oct-93 AUTHOR: User Contributed REVISION: 1.4 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 39 NUMBER OF LINES: 115 CONFIGURATION: Fluke 5700A ============================================================================= * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Source: Service Manual PM P/N: @ * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * TUR Configuration: * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * When the T.U.R. is < 2:1, the source instrument, e.g. 5700A, must be * characterized and appropriate corrections entered in the correction file * (e.g. 5700.COR). Where T.U.R.'s are > 2:1 but < 4:1, the test tolerance * is reduced by root sum square (rss) techniques. * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * This procedure is not executable in MET/CAL versions below ver. 2.00 * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Fluke make no warranty, expressed or implied, as to the fitness or * useability of this procedure. * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Main Procedure TESTS: Auto Calibration * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * The following subprocedures are called from this main procedure: * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DRAW 7 78 7 78 75 5 50 55 1.002 DISP Turn POWER on. [DRAW10,5,F] 1.003 DISP Allow the UUT to warm up for 30 min. and proceed the 1.003 DISP test at an ambient temp. of 23 +/- 1 deg. Celsius. 1.004 ASK- R 1.005 DISP Shortcircuit the 0 and the V-Ohm terminals to check 1.005 DISP the zero point (000.00 mV +/- 1 digit). 1.005 DISP Readjust the zero point if necessary 1.006 DISP Remove the shortcircuit, and connect testleads to the 1.006 DISP 0 and the V-Ohm terminals. 1.007 HEAD {DC VOLTAGE TEST} 1.008 DISP Set meter in range 0.2 1.009 5700 200 190mV 0.1/ 0.1% 2W 2.001 5700 200 -190mV 0.1/ 0.1% 2W 3.001 DISP Set meter in range 2 3.002 5700 2 1.9V 0.1/ 0.1% 2W 4.001 5700 2 -1.9V 0.1/ 0.1% 2W 5.001 DISP Set meter in range 20 5.002 5700 20 19V 0.1/ 0.1% 2W 6.001 5700 20 -19V 0.1/ 0.1% 2W 7.001 DISP Set meter in range 200 7.002 5700 200 190V 0.1/ 0.1% 2W 8.001 5700 200 -190V 0.1/ 0.1% 2W 9.001 DISP Set meter in range 2000 9.002 5700 2000 1000V 0.4/ 0.2% 2W 10.001 5700 2000 -1000V 0.4/ 0.2% 2W 11.001 HEAD {AC VOLTAGE TEST} 11.002 DISP Set PM2522 in Vac range. 11.002 DISP Set meter in range 0.2 11.003 5700 200 190mV 0.3/ 0.3% 100H 2W 12.001 5700 200 190mV 0.3/ 0.3% 10kH 2W 13.001 DISP Set meter in range 2 13.002 5700 2 1.9V 0.3/ 0.3% 100H 2W 14.001 5700 2 1.9V 0.3/ 0.3% 10kH 2W 15.001 DISP Set meter in range 20 15.002 5700 20 19V 0.3/ 0.3% 100H 2W 16.001 5700 20 19V 0.3/ 0.3% 10kH 2W 17.001 DISP Set meter inrange 200 17.002 5700 200 190V 0.3/ 0.3% 100H 2W 18.001 5700 200 190V 0.3/ 0.3% 10kH 2W 19.001 DISP Set meter in range 2000 19.002 5700 2000 600V 0.6/ 0.5% 100H 2W 20.001 HEAD {DIRECT CURRENT TEST} 20.002 DISP Disconnect testlead from V-Ohm terminal and connect 20.002 DISP it to the Ampere terminal. 20.002 DISP Set PM2522 in IDC range. 20.003 DISP Set meter in range 0.2 20.004 5700 200 190uA 0.25/ 0.25% 2W 21.001 5700 200 -190uA 0.25/ 0.25% 2W 22.001 DISP Set meter in range 2 22.002 5700 2 1.9mA 0.25/ 0.25% 2W 23.001 5700 2 -1.9mA 0.25/ 0.25% 2W 24.001 DISP Set meterin range 20 24.002 5700 20 19mA 0.25/ 0.25% 2W 25.001 5700 20 -19mA 0.25/ 0.25% 2W 26.001 DISP Set meter in range 200 26.002 5700 200 190mA 0.25/ 0.25% 2W 27.001 5700 200 -190mA 0.25/ 0.25% 2W 28.001 DISP Set meter in range 2000 28.002 5700 2 1.9A 0.25/ 0.25% 2W 29.001 5700 2 -1.9A 0.25/ 0.25% 2W 30.001 HEAD {ALTERNATING CURRENT TEST} 30.002 DISP Set PM2522 in IAC range. 30.003 5700 2 1.9A 0.25/ 0.25% 100H 2W 31.001 DISP Set meter in range 200 31.002 5700 200 190mA 0.25/ 0.25% 100H 2W 32.001 5700 200 190mA 0.25/ 0.25% 1kH 2W 33.001 DISP Disconnect testlead from Ampere terminal and connect 33.001 DISP it to the V-Ohm terminal. 33.002 HEAD {RESISTANCE TEST} 33.003 DISP Set PM2522 in kOhm range. 33.003 DISP Set meter in range 0.2 33.004 5700 200 190Z 0.2/ 0.3% 2W 34.001 DISP Set meter in range 2 34.002 5700 2 1.9kZ 0.2/ 0.3% 2W 35.001 DISP Set meter in range 20 35.002 5700 20 19kZ 0.2/ 0.3% 2W 36.001 DISP Set meter in range 200 36.002 5700 200 190kZ 0.2/ 0.3% 2W 37.001 DISP Set meter in range 2000 37.002 5700 2 1.9MZ 0.2/ 0.3% 2W 38.001 DISP Set meter in range 20Mohm 38.002 5700 20 19MZ 0.2/ 0.3% 2W 39.001 END